Light scatter
Last updated
Was this helpful?
Last updated
Was this helpful?
If a light scatter parameter is not being calibrated, proceed to processing .fcs files by clicking 'Next'.
If a light scatter parameter is being calibrated, click the '+' icon. Currently, only side scatter calibration is supported. Flow cytometer light scatter calibrations supported can be found here
From the 'Scatter Parameter' dropdown select the light scatter parameter that you wish to calibrate.
Alter the 'Scatter Wavelength (nm)' field to the wavelength that light scatter calibration is being performed at. Typically this is either 405 or 488 nm.
From the 'Bead Set' dropdown, select the calibration beads that were used as calibrators for the acquired data.
The median intensity of each bead should then be inputted into the 'Acquired Stat (au)' field. 'Acquired CV (%)' are optional and only used for illustrative purposes on the outputted scatter-diameter plots.
Further advanced light scatter calibration parameters can be applied or modified by clicking the 'Advanced Settings' button (No.4). Otherwise, proceed to processing .fcs files
Core-shell models for extracellular vesicles are added as default. The refractive index for their core and shell of the particle can be modified, along with the 'Shell Thickness (nm)'. How the EV models are named within the .fcs file can also be modified by editing the 'Parameter Name' fields.
Additional core-shell models can be added by click the '+' icon or removed by selecting the model and click the '-' icon.
Solid sphere models are not applied by default, but can be added by clicking the '+' icon under the 'Solid Sphere Parameters' table.
Sometimes it can be useful to see where a population of know sizes appears within the outputted scatter-diameter curve to have an idea of it's effective refractive index e.g. viruses or beads.
Plot datapoints can be specific by clicking the 'Particle Datapoints' tab and adding a datapoint by selecting the '+' icon.
The diameter and acquired statistic are required for plotting a datapoint with the CVs being optional.
The population name under the legend can be modified by editing the 'Name' field.
Modelling parameters relating the collection angle and geometry of the flow cytometer can be customized by selecting the 'Modelling Parameters' tab.
The default x-axis and y-axis limits for scatter QC plots is fixed and may be sub-optimal to illustrate data in some cases. These limits can be modifed by selecting the 'Plot Output Settings' tab.
An example of performing light scatter calibration with a pre-determined half-angle can be found .